A simple direct RT-LAMP SARS-CoV-2 saliva diagnostic

Michael J. Flynn, Olga Snitser, James Flynn, Samantha Green, I. Yelin, M. Szwarcwort-Cohen, R. Kishony, M. Elowitz

medRxiv, 2020

Widespread, frequent testing is essential for curbing the ongoing COVID-19 pandemic. Because its simplicity makes it ideal for widely distributed, high throughput testing, RT-LAMP provides an attractive alternative to RT-qPCR. However, most RT-LAMP protocols require the purification of RNA, a complex and low-throughput bottleneck that has often been subject to reagent supply shortages. Here, we report an optimized RT-LAMP-based SARS-CoV-2 diagnostic protocol for saliva and swab samples. In the protocol we replace RNA purification with a simple sample preparation step using a widely available chelating agent, as well as optimize key protocol parameters. When tested on clinical swab and saliva samples, this assay achieves a limit of detection of 105 viral genomes per ml, with sensitivity close to 90% and specificity close to 100%, and takes 45 minutes from sample collection to result, making it well suited for a COVID-19 surveillance program.

Cited by 9 publications.

Field of study: Computer Science

10.1101/2020.11.19.20234948